Vacuum for Concentrators

Requirements:

  • chemical resistance
  • tolerance to condensates
  • appropriate ultimate vacuum (depending on solvent)

Notices for Pumpchoosing:

Ultimate vacuum down to 7 mbar: For organic solvents and/or small amounts of water
Ultimate vacuum down to 1.5 mbar: For water-based samples and low temperatures, e.g.
DMF for T > 20 °C
DMSO for T > 50 °C
Ultimate vacuum
10-1 mbar:
For quick “drying”, low temperature
Ultimate vacuum
10-3 mbar:
For high-boiling-point solvents to be evaporated at low temperature; drying with minimal residues

Pumping unit equipment Ultimate vacuum
                                       down to 7 mbar down to 1,5 mbar

down to 0.5 mbar   

< 3x10-1 mbar < 10-3 mbar

Without flow control

MZ 2C NT
+AK+EK

 

MD 1C
+AK+EK
MD 4C
+AK+EK

PC 8 MV 10C 

 

RP 12C
RP 15C
RP 22C
RP 35C

PC 8 RC 6 

With manual flow control

PC 101 NT

 

PC 201 NT

 

 

 

 

With 2-point vacuum control

PC 510 NT
PC 511 NT
PC 520 NT

PC 610 NT
PC 611 NT
PC 620 NT

 

 

 

Continuous vacuum control

PC 3002 VARIO

PC 3001
PC 3004
PC 3012 VARIO

PC 3003 VARIO
PC 3010 VARIO