DCP 3000

Description

Vacuum Gauge DCP 3000 with Pressure Transducer VSK 3000

Vacuum measurement in the range of 1080 - 0,1 mbar (hPa)

  • precise
  • robust
  • for variable use
  • excellent chemical resistance, gas type independent

The versatile and fully electronic vacuum gauge DCP 3000 with the external capacitive pressure transducer VSK 3000 is made for vacuum measurement and process monitoring. The communication of the DCP 3000 with external components is done via VACUU∙BUS control allowing automatic component recognition. Thus up to four pressure transducers can be displayed. The DCP 3000 can often replace gas-type dependent and non corrosion resistant vacuum gauges (e.g. Pirani type).  

Advantages

  • Displays over 4 orders of magnitude (Atm to 0.1 mbar)

  • Pressure transducer with excellent chemical resistance

  • gas type independent measurement

  • Extra high long-term and temperature stability

  • User-selectable pressure units (Torr, mbar, hPa), digital calibration

  • On-the-spot measuring combined with network capability

  • Up to four external gauge heads at distances up to 30 m (see accessories)
  • Venting capability (accessory)
  • Serial interface RS 232C for data communication

  • Also for rough conditions with splash-proof VSK 3000 

Technical data

 

Measuring range 1080–0.1 mbar (hPa), 810–0.1 Torr
Uncertainty of measurement < ±1 mbar (0.75 Torr) ±1 digit
Temperature coefficient           < 0.07 mbar/K
Gauge head VSK 3000 / pressure transducer made of alumina ceramic, cable 2.0 m
Vacuum connection small flange NW 16 with screw fitting for PTFE tubes 10/8 mm; stepped screw-in hose nozzle NW 6/10 (included)
Protection rating IEC 529

Gauge head: IP 54 (against splash water)

Display unit front side: IP 42 (against dripping water)

Mains supply

90 - 264 V~ / 47 - 63 Hz

Dimensions (L x W x H) 

138 x 124 x 115 mm (display unit DCP 3000)

D60 x 60 (pressure transducer VSK 3000)

Weight vacuum gauge 0.44 kg
Weight pressure transducer 0.18 kg

Order data

 

DCP 3000 with external gauge head VSK 3000

lead, ready for use

Plug  Cat. No.
Vacuum gauge DCP 3000 with VSK 3000 CEE/CH/UK/US/AU 68 31 70
DKD calibration on first delivery 90 02 14
DKD recalibration 90 02 15
Accessories:
External gauge head VSK 3000 VACUU•BUS 63 66 57
Air admittance valve VBM-B with KF 16 NW VACUU•BUS 67 42 17
VACUU•BUS extension cable, 2 m VACUU•BUS 61 25 52
Y adapter VACUU•BUS VACUU•BUS 63 66 56