DVR 2
Description
Digital / Analogue Vacuum Gauge DVR 2
Vacuum measurement in the range of 1080 - 1 mbar (hPa)
- Large analogue and digital vacuum display with
clear and precise readout - mains-independent operation for variable use
- excellent accuracy
The DVR 2 is a fully electronic, versatile vacuum gauge for the measuring range from atmospheric pressure to 1 mbar. The DVR 2 has an integrated capacitive transducer made of alumina ceramic with outstanding corrosion resistance and superior long-term stability. A particular advantage of the DVR 2 is its mains-independent battery operation. Its simple operation and the combined digital/analogue vacuum reading for quick trend detection and precise readout make the DVR 2 very versatile. It meets high requirements concerning accuracy, safety and environmental protection.
Advantages
- Large analogue and digital vacuum display: quick trend detection, clear and precise readout
- integrated capacitive alumina-ceramic transducer: very good corrosion resistance and long-term stability
- fully electronic – no fragile mechanics
- User-selectable pressure units (Torr, mbar, hPa)
- Digital calibration
- EMC-compatible
- Mains-independent operation
Technical data
| Measuring range | 1080–1 mbar (hPa), 810–1 Torr |
| Pressure readout | digital and analog |
| Uncertainty of measurement | < ±1 mbar (0.75 Torr) ±1 digit |
| Principle of measurement | capacitive, gas-type independent absolute measuring |
| Transducer | integrated, made of alumina ceramic |
| Vacuum connection | small flange NW 16 and stepped screw-in hose nozzle NW 6/10 |
| Temperature coefficient | < 0.07 mbar/K |
| Housing | robust plastic housing (PP) with high chemical resistance |
| Power supply unit | 9 V lithium battery, 1.2 Ah Ultralife U9 VL |
| Dimensions (L x W x H) | 115 x 115 x 66 mm |
| Weight | 0.40 kg |
Order data
Vacuum gauge DVR 2 | Cat. No. |
|---|---|
| DVR 2 | 68 29 02 |
| Items supplied: Vacuum gauge with on/off switch, 9 V battery, mounting rod and instructions for use. | |
| Battery 9 V Lithium 1.2 Ah | 61 22 20 |
| DKD calibration on first delivery | 90 02 16 |
| DKD recalibration | 90 02 17 |