VNC 1

Description

Vacuum Network Controller VNC 1

Vacuum measurement and control in the range of
1100 - 1 mbar (hPa)

The Vacuum Network Controller VNC 1 provides vacuum measurement and control within the range from 1100 to 1 mbar. It operates with vacuum pumps or valves. Particularly with VARIO and VARIO-SP vacuum pumps, precise vacuum process control can be achieved. A cooling water valve can be operated too, e.g. for an exhaust vapour condenser. The parts of the gauge head in contact with media are made of stainless steel and fluoroelastomer, the pressure transducer is made of alumina ceramic. The VNC 1 is also available as a built-in version for integration into lab furniture.

The Vacuum Network Controller VNC 1 is especially optimized for application dependent control of vacuum pumps in local vacuum networks.


There are five modes of operation:

Pressure control
Two set point control of a vacuum pump, a vacuum line valve or pinpoint-precision vacuum control with a VARIO Diaphragm Pump

Continuous pumping
Control of a vacuum pump or motor speed control (VARIO)

VACUU•LAN®
Control of a VARIO Diaphragm Pump and/or on/off control of a vacuum line valve or a vacuum pump, according to a specified vacuum level or timing, for fully automatic control of a vacuum network

Monitoring
Control of a vacuum pump or a vacuum line valve according to specifications for min./max. pressure and shutdown time. Typical processes: aspiration, filtration, vacuum drying

Process management of Chemistry Hybrid Pump RC 6
monitoring of internal vacuum of the RC 6, optionally with vacuum valve and bypass valve

Advantages

  • Precise and stable pressure transducer with high chemical resistance
  • very compact design
  • easy menu-guided operation
  • flexible parameter adjustment for vacuum networks
  • acoustic and visual alert signals
  • interactive communication via RS 232C interface

    Technical data

     

    Measurement range1100-1 mbar (hPa), 825-1 Torr
    Display alphanumeric LCD display
    Max. range of pressure control 1060–1 mbar (hPa), 795–1 Torr (depending on gas load and pump)
    Uncertainty of measurement< ±1 mbar (0.75 Torr), ±1 digit
    Temperature coefficient of transducer< ±0.07 mbar/K
    Pressure transducer capacitive, made of alumina ceramic, gas type independent 
    Vacuum connection G 1/4", screw-in hose nozzle NW 6/10
    Interface RS 232C
    Power consumption 3 VA (idle), 10 VA (max.)
    Mains supply 100-230 V ~ ±10% 50-60 Hz
    Dimensions (L x W x H) benchtop / built-in version* 144 x 90 x 68 mm / 135 x 90 x 61 mm
    Weight 0.8 kg
    * versions including front panel on request

    Order data

     

    Vacuum network controller VNC 1
    with on/off switch, IEC socket and instructions for use.
    Best.-Nr.
    VNC 168 29 70
    VNC 1 panel-mountable built-in version (without panel)* 68 29 75
    Electromagn. valve VV 6C, small flange NW 16, hose nozzle NW 6/10 67 40 91
    Cooling water valve 24 V =, G 3/4", G 1/2", hose nozzle DN 6 67 60 13
    Adapter cable for VARIO pumps/cooling water valve 63 71 76
    DKD calibration on first delivery 90 02 16
    DKD recalibration 90 02 17