VNC 2
Description
Vacuum Network Controller VNC 2
Vacuum measurement and control in the range of
1100 - 1 mbar (hPa)
The VNC 2 enables an automatic, on-demand vacuum control either via 2-set-point valve control or switching on/off the pump. The integrated pressure transducer is chemical resistant and offers an outstanding measuring accuracy independent of gas type. The device is designed for precise and reliable control of recurring routine jobs. The vacuum controller VNC 2 can only be operated with components compatiblle with the VACUU·BUS-system. Special versions of the controller, e. g. for installation in laboratory furniture, are available. Further information on request.
The Vacuum Network Controller VNC 2 is especially optimized for application dependent control of vacuum pumps in local vacuum networks and serves as controller in the network pumping units PC 500 / 600 LAN NT.
There are five modes of operation:
Pressure control
Two set point control of a vacuum pump or a solenoid-operated vacuum line valve
Continuous pumping
Control of a vacuum pump
VACUU•LAN®
Control of a VARIO Diaphragm Pump and/or on/off control of a solenoid-operated vacuum line valve or a vacuum pump, according to a specified vacuum level or timing, for fully automatic control of a vacuum network
Monitoring
Control of a vacuum pump or a solenoid-operated vacuum line valve according to specifications for min./max. pressure and shutdown time. Typical processes: aspiration, filtration, vacuum drying
Process management of Chemistry Hybrid Pump RC 6
monitoring of internal vacuum of the RC 6, optionally with vacuum valve and bypass valve
Advantages
- Precise and stable pressure transducer with high chemical resistance
- very compact design
- easy menu-guided operation
- flexible parameter adjustment for vacuum networks
- acoustic and visual alert signals
- interactive communication via RS 232C interface
Technical data
| Measurement range | 1100-1 mbar (hPa), 825-1 Torr |
| Display | alphanumeric LCD display |
| Max. range of pressure control | 1060–1 mbar (hPa), 795–1 Torr (depending on gas load and pump) |
| Uncertainty of measurement | < ±1 mbar (0.75 Torr), ±1 digit |
| Temperature coefficient of transducer | < ±0.07 mbar/K |
| Pressure transducer | capacitive, made of alumina ceramic, gas type independent |
| Vacuum connection | PTFE-tubing conn. 10/8 mm with hose nozzle DN 6/10 mm (small flange KF DN 16 opt.) |
| Interface | RS 232C |
| Power consumption | 7.5 W |
| Mains supply | 100-230 V ~ 50-60 Hz |
| Dimensions (L x W x H) benchtop / built-in version* | 163 x 90 x 68 mm / 138 x 90 x 120 mm |
| Weight | 0.8 kg |
| * versions including front panel on request | |
Order data
| Vacuum network controller VNC 2 with on/off switch, IEC socket and instructions for use. | Best.-Nr. |
|---|---|
| VNC 2 | 68 30 70 |
| VNC 2 panel-mountable built-in version with valve (without panel)* | 68 30 90 |
| Mains cable please order seperately: | |
| Mains cable CEE | 61 20 58 |
| Mains cable CH | 67 60 21 |
| Mains cable UK | 67 60 20 |
| Mains cable US | 61 20 65 |
| Electromagnetic valve VV-B 6C, small flange KF DN 16, hose nozzle DN 6/10 | 67 42 91 |
| Cooling water valve VKW-B, G 3/4", G 1/2", hose nozzle DN 6 | 67 42 20 |
| Extension cable VACUU·BUS, 2 m | 61 25 52 |
| DKD calibration on first delivery | 90 02 16 |
| DKD recalibration | 90 02 17 |
* versions including front panel on request