VNC 2

Description

VNC 2

The vacuum controller VNC 2 makes automatic, on-demand vacuum control possible either via two-point valve control or by switching the pump on/off. The version VNC 2 VARIO offers continuous vacuum control with VARIO® pumps. The integrated alumina ceramic diaphragm vacuum sensor is chemically resistant and offers outstanding measuring accuracy independent of gas type. The VNC 2 vacuum controller is designed for precise and reliable control of recurrent routine jobs that do not require regular adjustment. It works with components compatible with the VACUU·BUS® system. Special versions of the controller, e.g., for mounting into laboratory furniture are available. Further information on request.

Performance features

  • on-demand control of process vacuum, cooling water and venting; special program for operating vacuum networks like VACUU·LAN®
  • easy changing of vacuum set points also during operation
  • automatic or manual setting of the pressure switching interval (hysteresis)
  • integrated relay for switching pumps on/off (except VNC 2 VARIO)
  • selectable acoustic alert signals

Technical data / Ordering data

Technical data

Technical dataUnit VNC 2
Vacuum sensorIntegr.
Upper measuring limitmbar/torr1100 / 825
Lower measuring limitmbar/torr1 / 1
Measurement principleCeramic diaphragm (alumina), capacitive, gas indep., absolute pressure
Measurement uncertainty< +- 1 mbar/hPa/torr / +- 1 digit (after adjustment, constant temp.)
Temperature coefficient< 0.07 mbar/hPa/0.05 torr /K
Ambient temperature range (operation)°C10 - 40
Ambient temperature range (storage)°C-10 - 60
Max. media temp. for continuous operation / short times°C40 / 80
Material of outer housingCoated metal housing
Degree of protectionIP 20
Max. power for VACUU·BUS® (24V DC)400 mA (line voltage: 110-230 V) / 340 mA (line voltage: 100-110 V)
InterfaceRS 232C
Dimensions (L x W x H)mm163 x 90 x 68
Weightkg0.8
ATEX-approvalII 3G IIC T3 X Internal Atm. only
Items suppliedVacuum controller complete with integrated vacuum sensor, ready for use, with manual.
AccessoriesRubber vacuum tubing DN 10 mm (686002)
PTFE tubing DN 10/8 mm (638644)
DAkkS calibration with first delivery (900216)
DAkkS recalibration (900217)
VACUU·BUS® accessories

VNC 2 - Dimension sheet

-Technical data are subject to change without notice-

Ordering data

Product nameMains voltage and frequencyPlugOrder no.
VNC 2 100-230 V~ 50-60 Hz
IEC plug EN 60320 683070

Product comparison

Product comparison

CVC 3000 --0.1 --
VNC 2 --1 --
CVC 3000 +
VSP 3000, KF DN 16
---1 x 10-3 ---
CVC 3000 +
VSP 3000, KF DN 25
---1 x 10-3 ---