DVR 3
Note: Your request will be forwarded immediately to our partner next to you

DVR 3

VACUUBRAND682903ca. 66 mmca. 116 mmca. 0.43 kg

The DVR 3 is a fully electronic vacuum gauge with ATEX category 2 approval for use in areas with potentially explosive atmospheres (measured gas and environment of gauge). The DVR 3 has an integrated ceramic diaphragm vacuum sensor for the range from atmospheric pressure to 1 mbar with capacitive readout. The sensor measures independently of the type of gas and is exceptionally resistant to chemicals. It features outstanding measuring accuracy, very low temperature depencence and excellent long-term stability. A particular advantage of the DVR 3 is its cordless, battery-powered operation. Its simple operation and the combined digital / analog vacuum reading for precise readout and quick trend detection, respectively, make the DVR 3 very versatile.

Performance features

  • large analog and digital vacuum display
  • mains independent power supply unit BVE 9V, with ATEX approval, mercury and cadmium free
  • enhanced battery life due to automatic shutdown and variable sampling rate
  • high reliability, low EMI emission level near detection limit, high degree of immunity to electromagnetic interference for use in industrial environments
  • capacitive alumina ceramic diaphragm vacuum sensor with excellent chemical resistance, accuracy and long term stability

Technical data

Unit Converter Interested? Send your inquiry now!
Technical dataUnit DVR 3
Upper measuring limitmbar/torr1080 / 810
Lower measuring limitmbar/torr1 / 1
Measurement principleCeramic diaphragm (alumina), capacitive, gas indep., absolute pressure
Accuracy of measurement< +- 1 mbar/hPa/torr / +- 1 digit (after adjustment, constant temp.)
Temperature coefficient< 0.07 mbar/hPa/0.05 torr /K
Ambient temperature range (operation)°C10 - 40
Ambient temperature range (storage)°C-10 - 60
Max. media temp. for continuous operation / short times°C40 / 80
Automatic switch-offUser-selectable: 1-1000 min (default 15 min) or continuous operation
Measurement cycleUser-selectable: Automatic or 1 x per 3s, 1 x per 1s, 3 x per 1s
Material of outer housingRobust plastic housing (conductive) with good chemical resistance
Degree of protectionIP 40
Power supplyPower supply unit BVE 9 V, ATEX - certification II 2G Ex ia IIC T5
Typical battery lifetime8.000 h, stand-by: 40.000 h
Dimensions (L x W x H)mm116 x 116 x 66
Weightkg0.43
ATEX-approvalII 2G Ex ia IIC T4
Items suppliedVacuum gauge complete with integrated vacuum sensor and power supply unit, ready for use, with manual.
AccessoriesPTFE tubing KF DN 16 (1000 mm: 686031)
Stainless steel tubing KF DN 16 (1000 mm: 673336)
Power supply unit BVE 9 V (637986)
DAkkS calibration with first delivery (900214)
DAkkS recalibration (900215)

DVR 3 - Dimension sheet

DVR 3 - Dimension sheet

-Technical data are subject to change without notice-

Ordering data

Product nameMains voltage and frequencyPlugOrder no.
DVR 3 9 V= -682903