Cold traps
At inlet pressures below 1 mbar cold traps using dry ice or liquid nitrogen as coolant provide important protection for your pump by separating condensates and aggressive media. Cold traps using liquid nitrogen may be essential with very volatile solvents and reduce the back streaming of oil molecules into the vacuum system. Furthermore a cold trap will considerably increase the effective pumping speed for vapors.
Particles reduce the lifetime of the oil and increase service demands. The full-flow oil filters efficiently filter micro particles out of the oil. The service indicator helps to avoid unnecessary filter changes.
![]() | ![]() |
SKF H 25 | GKF 1000i |
PERFORMANCE FEATURES
Cold trap SKF
- sturdy, easy to clean
- easy to disassemble
- two-wall design with good conductance
- long operating time per coolant filling
- easy condensate drainage without disassembling
Cold trap GKF
- highly reflective insulation jacket for extended coolant life
- vertical window: Direct observation of condensate and coolant levels
- PTFE stopcock: Drain condensate without disassembly
- sheet metal shield for protection against external damage and implosion
TECHNICAL DATA | SKF H 25 | SKF H 40 | GKF 1000i | |
Inlet | small flange KF DN 25 | Small flange KF DN 40 | NS 29/32 female ground joint | |
Outlet | small flange KF DN 25 | Small flange KF DN 40 | Glass tube 22 mm | |
Materials | Stainless steel, FKM, NBR | Stainless steel, FKM, NBR | Borosilicate glass, PTFE, FKM | |
Volume catchpot | ml | 500 | 500 | 250 |
Coolant capacity | ml | 1000 | 1000 | 1000 |
Typ. coolant life | h | 12 h | 12 h | 14 h |
Dimensions (L x W x H), approx. | mm | 166 x 136 x 303 | 166 x 140 x 319 | D 148 x 580 |
ORDERING INFORMATION | |
SKF H 25 | 20667051 |
SKF H 40 | 20667053 |
GKF 1000i | 20667056 |