Vacuum gauge Set DCP 3000 + MPT

Description

VACUUBRAND20683175ca. 114 mmca. 124 mmca. 0.44 kg20683175

For the range from atmospheric pressure down to high vacuum (5 x 10-9 mbar) the vacuum gauge DCP 3000 is available with the combined thermal conductivity / cold cathode (Pirani / Penning) vacuum sensor MPT.

Performance features

  • external gauge head for measuring on the spot
  • wide measurement range from atmospheric pressure to high vacuum (5 x 10-9 mbar) due to combined thermal conductivity / cold cathode measurement system (Pirani / Penning)
go to top

Technical data

Unit Converter
Technical dataUnit Set DCP 3000 + MPT
Vacuum sensorMPT 200
Cable length of vacuum sensorm2
Upper measuring limitmbar/torr1 x 103 / 7.5 x 102
Lower measuring limitmbar/torr5 x 10-9 / 3.7 x 10-9
Measurement principleCombined thermal conductivity (Pirani) / cold cathode (Penning)
Accuracy of measurement1 x 10-8 - 2 x 10-3 mbar: +- 25%; 2 x 10-3 -10 mbar: +- 10%; 10 - 100 mbar: +- 30%; 100 - 1000 mbar:+- 50%
Ambient temperature range (operation)°C10 - 40
Ambient temperature range (storage)°C-10 - 60
Max. media temp. for continuous operation / short times°C40 / 80
Material of outer housingRobust plastic housing with good chemical resistance
Degree of protectionIP 20
Degree of protection of front side (also built-in version)IP 42
Max. power for VACUU·BUS® (24V DC)4 A
Cable length of external power supplym2
InterfaceRS 232C
Dimensions (L x W x H)mm144 x 124 x 114
Weightkg0.44
Items suppliedVacuum gauge with external vacuum sensor MPT, ready for operation, with wall-plug power supply and manual.
AccessoriesStainless steel tubing KF DN 25 (1000 mm: 20673337)
Vent valve VB M-B (20674217)
Vacuum sensor VSK 3000 (20640530)
Vacuum sensor VSP 3000 (20636163)
Vacuum sensor MPT 200 (20683177)

-Technical data are subject to change without notice-

Ordering data

Product nameMains voltage and frequencyPlug* Order no.
Set DCP 3000 + MPT 100-230 V~ 50-60 Hz
CEE/CH/UK/US/AUS/CN 20683175